V10 Air Plasma Sterilizer
The V10 Air Plasma Sterilizer is a low temperature sterilzer.
Sterilization Cycle Specifications:
- Type of sterilization cycle: Single cycle sterilizes most general surgical and laparoscopic instruments*
- Sterilization process: Air Plasma SterilizationTM process, terminal sterilization using overkill method to provide SAL of 10-6
- Total sterilization cycle time: Approximately 4 hours
- Preconditioning time: 15-20 min+
- Sterilizing time: 3 hours
- Ventilating time: 45 minutes
- Sterilant: Reactive oxygen and nitrogen species, primarily ozone, nitrogen dioxide, and hydroxyl radicals
- Sterilant production: Sterilant gases produced from preconditioned air along the surface of the plasma sheet
- Sterilant delivery: Sterilant gases produced inside sterilization chamber
- Sterilant removal: Sterilant gases are converted back into air by pumping through Carulite® 200 granular catalyst
- Critical process parameters: Air temperature, pressure, and humidity are monitored during preconditioning, voltage feedback from plasma production is monitored throughout the sterilizing step
- Sterility assurance: Biological indicators (≥106 spores of Geobacillus stearothermophilus)
Technical Specifications:
- Exterior dimensions 28.75 in x 19.25 in x 18.75 in (73 cm x 49 cm x 48 cm)
- Installation space requirement 30.75in x 21.25 in x 20.75 in (2 inches on all sides)
- Sterilization chamber dimensions 18 in x 16 in x 7 in (46 cm x 41 cm x 18 cm)
- Sterilization chamber volume 33 L (8.7 gal)
- Sterilization tray dimensions 16 in x 10 in x 4 in (41 cm x 25 cm x 10 cm)
- Sterilization tray volume 10.5 L (2.8 gal)
- Weight 110 lbs (49.9 kg)
- Operating temperature 5-40 °C (41-104 °F)
- Storage temperature -25-60 °C (-13-140 °F)
- Electrical requirements 110-120 VAC, 50/60Hz, 6.5 A max (250 V 8 A rated fuse)
- Operating relative humidity 0%-95%
- Operating altitude Up to 2,000 m (6,562 ft), 545-775 Torr (72.4-103.4 kPa)
1 Year Warranty